Dot-x SERIES Specifiche Pagina 48

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 47
DOTX-45 Dissolved Oxygen System Part 7 –Calibration
DOTX-45 47 -
4. Press the ENTER key. The screen will display a flashing 1-spl for 1-point, a
%-sat for Saturation calibration, or a 3-zer for zero calibration. Using the UP
arrow key, set for a 1-spl calibration and press ENTER.
5. The system now begins acquiring data for the calibration value. As data is
gathered, the units for ppm and temperature may flash. Flashing units
indicate that this parameter is unstable. The calibration data point acquisition
will stop only when the data remains stable for a pre-determined amount of
time. This can be overridden by pressing ENTER.
6. If the data remains unstable for 10 minutes, the calibration will fail and the
message Cal Unstable will be displayed.
7. The screen will display the last measured ppm value and a message will be
displayed prompting the user for the lab value. The user must then modify the
screen value with the arrow keys and press ENTER. The system then
performs the proper checks.
8. If accepted, the screen will display the message PASS with the new slope
reading, then it will return to the main measurement display. If the calibration
fails, a message indicating the cause of the failure will be displayed and the
FAIL icon will be turned on.
7.13 Dissolved Oxygen Air Span Cal (% sat)
This is the recommended method for air temperatures greater than about 10°C;
however, it requires that the sensor be removed from the process and cleaned.
Once cleaned, the sensor is held in air and allowed time to adjust to the air
temperature. As the sensor temperature equilibrates, the transmitter
automatically calculates the new 100% saturation point utilizing the temperature
readings and the barometric pressure user data located on page 64. This method
therefore requires no user input during calibration. Note: It is very important to
allow enough time for the sensor to completely temperature equilibrate
with the surrounding air.
This method requires that the sensor be removed from the process, cleaned, and
Covered, shielding the sensor from direct sunlight. The sensor membrane must
be dry for this procedure, and not submerged in liquid. This method requires no
user input during calibration; however, if this is the first time the system is being
installed and calibrated, make sure to enter the proper atmospheric pressure
data and process conductivity data prior to calibration.
1. Remove the sensor from the process. Clean and rinse if necessary with
water, paying particular attention to cleaning the membrane.
Vedere la pagina 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 68 69

Commenti su questo manuale

Nessun commento