DOTX-45 Dissolved Oxygen System Part 6 – Configuration
DOTX-45 40 -
delay setting for the controller. The entry value is limited to a
value between 0 and 999 seconds. Press ENTER to initiate
user entry mode, and the value will flash. Use arrow keys to
modify value; range is 0 to 999 seconds. Press ENTER to
store the new value.
*A Phasing This function establishes the direction of the relay trip.
When phase is HI, the relay operates in a direct mode.
Therefore, the relay energizes and the LCD indicator
illuminates when the oxygen value exceeds the setpoint.
When the phase is LO, the relay energizes and the LCD
indicator illuminates when the oxygen level drops below the
setpoint. The failsafe setting does have an impact on this
logic. The description here assumes the failsafe setting is
OFF. Press ENTER to initiate user entry mode, and the
entire value will flash. Use the UP arrow key to modify the
desired value; selections include HI for direct operation or
LO for reverse operation. Press ENTER to store the new
value.
See Figure 6-3 below for a visual description of a typical
control relay application.
Figure 6-3 Control Relay Example, Hysteresis and Two opposite Phase Options
NOTE: If the temperature units are changed between °
and °
F (see Temp Units in this section), the default
setting for this output will change between 100 °
and 212 °F accordingly.
Setpoint: 1.000 ppm
Hyst: 0.050
Delay: 000
Failsafe: OFF
When value rises to ≥ 1.000 ppm, relay closes.
When value falls to ≤ 0.950 ppm, relay opens.
When value falls to ≤ 1.000 ppm, relay closes.
Settings:
}
1.000 ppm
0.950 ppm
PHASE: HI
HYSTERESIS
OR
“DEAD BAND”
X
OFF
OR
ON