DOTX-45 Dissolved Oxygen System Part 6 Configuration
DOTX-45 37 -
AC OPERATED UNITS ONLY
*Relay B Mode Relay B can be used in a number of ways: as a setpoint
control, as an alarm, or as the control logic for an automatic
sensor cleaning system. The settings for Relay B Mode are
CON or FAIL. CLn1 and CLn2 are not used with this
instrument.. The first two modes function identically to the
corresponding modes on Relay A.
Temp Units This function sets the display units for temperature
measurement. Press ENTER to initiate user entry mode,
and the entire value will flash. Use the UP arrow key to
modify the desired display value. The choices are °F and
°C. Press ENTER to store the new value.
6.25 Control Menu [CONTROL]
The Control Menu contains all of the output control user settings:
Set 4 mA
Set 20 mA
[Iout1=D.O.] These functions set the main 4 and 20 mA current loop
output points for the transmitter. The units displayed depend
on the selection made in the CONFIG menu for Iout #1
Mode. Also, when the Relay Option Board is installed, the
units will also display #1 or #2 since there are actually two
analog outputs present in this version.
The value stored for the 4 mA point may be higher or lower
than the value stored for the 20 mA point. The entry values
are limited to values within the range specified in Set
Range, and the 4 mA and the 20 mA point must be
separated by at least 1% of this range Use the LEFT arrow
key to select the first digit to be modified. Then use the UP
and LEFT arrow keys to select the desired numerical value.
Press ENTER to store the new value.
Set PID 0%
Set PID 100%
[Iout1=PID] If the PID is enabled, this function sets the minimum and
maximum controller end points. Unlike the standard 4-20
mA output, the controller does not scale output values
across the endpoints. Rather, the endpoints determine
where the controller would normally force minimum or
maximum output in an attempt to recover the setpoint (even